SYSTEM NOISE FLOOR VS. OPERATING FREQUENCY
Frequency
(MHz)
Noise Floor
(Electrons)
Dynamic Range*
(Bits)
Dynamic Range*
(dB)
El/ADU System Gain
1 40 11.49 69.17 29.2 2.5X
2 48 11.23 67.59 29.2 2.5X
3 50 11.17 67.23 29.2 2.5X
4 54 11.06 66.57 29.2 2.5X
5 56 11.00 66.25 29.2 2.5X
6 58 10.95 65.95 29.2 2.5X
Note: Dynamic Range Calculated using KAF1600 Sensor, Full Well = 115000 Electrons Measured
0
10
20
30
40
50
60
01234567
Frequency (MHz)
Noise Floor (Electrons)
10
10.4
10.8
11.2
11.6
12
01234567
Frequency (MHz)
Bits
Figure 15: Measured Performance: Dynamic Range